Ifm SB1257

Flow meter with integrated backflow prevention and display. SBG32KM0FRKG

Na prośbę. Prześlij nam wiadomość e-mail z zapytaniem o cenę i dostępność.
"ucnguBrctvdgcxgt0eqo"

Opis

Parametry techniczne

Process connection: threaded connection G 1 1/2 internal thread
Measuring range: -10...100 °C 14...212 °F
-10...100 °C: 14...212 °F
Special feature: Gold-plated contacts
Media: Liquids; oil
Note on media: oil with viscosity: 10 mm²/s (40 °C)
Medium temperature [°C]: -10...100
Pressure rating [bar]: 63
Note on pressure rating: at medium temperature >70°C: 50
Operating voltage [V]: 18...30 DC; (according to EN 50178 SELV/PELV)
Current consumption [mA]: < 50
Protection class: III
Reverse polarity protection: yes
Power-on delay time [s]: < 3
Total number of outputs: 2
Output signal: switching signal; analogue signal; frequency signal; IO-Link
Output function: normally open / normally closed; (parameterisable)
Max. voltage drop switching output DC [V]: 2
Max. current load per output [mA]: 150; (200: ...60 °C; Ambient temperature; 250: ...40 °C; Ambient temperature)
Analogue current output [mA]: 4...20
Max. load [Ω]: 500
Short-circuit protection: yes
Overload protection: yes
Frequency of the output [Hz]: 0...10000
Measuring range: 4...200 l/min 0.24...12 m³/h 65...3170 gph 1.1...52.8 gpm
4...200 l/min: 0.24...12 m³/h
Display range: 0...240 l/min 0...14.4 m³/h 0...3805 gph 0...63.4 gpm
0...240 l/min: 0...14.4 m³/h
Resolution: 0.1 l/min 0.01 m³/h 1 gph 0.1 gpm
0.1 l/min: 0.01 m³/h
Set point SP: 1.4...200 l/min 0.08...12 m³/h 20...3170 gph 0.3...52.8 gpm
1.4...200 l/min: 0.08...12 m³/h
Reset point rP: 0...198.6 l/min 0...11.92 m³/h 0...3150 gph 0...52.5 gpm
0...198.6 l/min: 0...11.92 m³/h
Frequency end point, FEP: 13.4...200 l/min 0.8...12 m³/h 210...3170 gph 3.5...52.8 gpm
13.4...200 l/min: 0.8...12 m³/h
In steps of: 0.02 l/min 0.02 m³/h 5 gph 0.1 gpm
0.02 l/min: 0.02 m³/h
Frequency at the end point FRP [Hz]: 10...10000
In steps of [Hz]: 10
Measuring dynamics: 1:50
In steps of: 10 Hz
Measuring range: -10...100 °C 14...212 °F
-10...100 °C: 14...212 °F
Display range: -32...122 °C -25.6...251.6 °F
-32...122 °C: -25.6...251.6 °F
Resolution: 0.1 °C 0.1 °F
0.1 °C: 0.1 °F
Set point SP: -9.3...100 °C 15.2...212 °F
-9.3...100 °C: 15.2...212 °F
Reset point rP: -10...99.3 °C 14...210.8 °F
-10...99.3 °C: 14...210.8 °F
In steps of: 0.1 °C 0.2 °F
0.1 °C: 0.2 °F
Frequency start point, FSP: -10...78 °C 14...172.4 °F
-10...78 °C: 14...172.4 °F
Frequency end point, FEP: 12...100 °C 53.6...212 °F
12...100 °C: 53.6...212 °F
Frequency at the end point FRP [Hz]: 10...10000
Accuracy (in the measuring range): ± 5 % MEW; (Q > 1 l/min; 20...70 °C Medium temperature)
Repeatability: ± 1 % MEW
Temperature drift: 0,029 °C / K
Accuracy [K]: 3 K (25°C; Q > 1 l/min)
Response time [s]: 0.01
Damping for the switching output dAP [s]: 0...5
In steps of [s]: 0.1
Damping for the analogue output dAA [s]: 0...5
In steps of [s]: 0.1
Dynamic response T05 / T09 [s]: T09 = 120 (Q > 1 l/min)
Parameter setting options: hysteresis / window; normally open / normally closed; switching logic; current/frequency output; damping for the switching output / analogue output; display can be rotated and switched off; standard unit of measurement; process value colour; calibration factor
Communication interface: IO-Link
Transmission type: COM2 (38,4 kBaud)
IO-Link revision: 1.1
SDCI standard: IEC 61131-9 CDV
IO-Link device ID: 1047 d / 0417 h
Profiles: Smart Sensor: Process Data Variable; Device Identification, Device Diagnosis
SIO mode: yes
Required master port type: A
Process data analogue: 2
Process data binary: 2
Min. process cycle time [ms]: 3.2
Ambient temperature [°C]: 0...60
Note on ambient temperature: medium temperature < 80 °C medium temperature < 100 °C: 0...40 °C
Storage temperature [°C]: -15...80
Protection: IP 65; IP 67
EMC: DIN EN 61000-6-2 DIN EN 61000-6-3
Shock resistance: DIN EN 60068-2-27 20 g (11 ms)
DIN EN 60068-2-27: 20 g (11 ms)
Vibration resistance: DIN EN 60068-2-6 5 g (10...2000 Hz)
DIN EN 60068-2-6: 5 g (10...2000 Hz)
MTTF [years]: 145
UL approval: UL Approval no. I007
UL Approval no.: I007
Pressure Equipment Directive: Sound engineering practice
Weight [g]: 2796.5
Materials: stainless steel (1.4404 / 316L); PBT+PC-GF30; PBT-GF20; PC; brass chemically nickel-plated
Materials (wetted parts): stainless steel (1.4401 / 316); stainless steel (1.4404 / 316L); brass (2.0371); brass chemically nickel-plated; PPS; O-ring: FKM
Process connection: threaded connection G 1 1/2 internal thread
Switching cycles mechanical: 10 million
Display: Display unit 6 x LED, green switching status 2 x LED, yellow measured values alphanumeric display, red/green alternating indication 4-digit programming alphanumeric display, 4-digit
Display unit: 6 x LED, green
switching status: 2 x LED, yellow
measured values: alphanumeric display, red/green alternating indication 4-digit
programming: alphanumeric display, 4-digit
Remarks: Recommendation: use a 200-micron filter. All data refer to oil with the following nominal viscosity: 10 mm²/s, 40 °C MW = measured value MEW = Final value of the measuring range
Pack quantity: 1 pcs.
Connection: Connector: 1 x M12; Contacts: gold-plated

Parametry logistyczne

ECCN:
Kraj pochodzenia: DE
EAN:
Waga: 2,80 kg
Długość: 0,00 m
Szerokość: 0,00 m
Wysokość: 0,00 m